risk-management-specialist — Claude Code skill: setup & guide
Implements ISO 14971:2019 risk management for medical devices across product lifecycles. It guides hazard analysis, 5x5 matrices, and risk control.
/plugin marketplace add alirezarezvani/claude-skills
/plugin install ra-qm-skills@claude-code-skillsWhat risk-management-specialist does
The skill establishes workflows for medical device lifecycle risk management per ISO 14971:2019. It guides planning, hazard identification, probability levels from P1 to P5, and severity scoring from S1 to S5. Checklist categories cover energy, biological, chemical, operational, and software hazards.
Workflows recommend specific methods: FMEA for components, FTA for system failures, HAZOP for process deviations, and Use Error Analysis. The instructions enforce EU MDR compliance, requiring risk reduction as far as possible (AFAP). Cost-benefit balancing under ALARP is excluded because EU notified bodies reject it.
Teams preparing medical software or hardware for regulatory review should install it. For consumer software without healthcare compliance, these workflows add unnecessary overhead.
What it helps with
- draft an ISO 14971 risk management plan for a medical device
- select appropriate risk analysis methods like FMEA, FTA, or HAZOP
- construct a 5x5 risk acceptability matrix
- verify risk reduction against EU MDR AFAP criteria
Example prompts for the agent
- Create an ISO 14971 risk management plan for an embedded medical device
- Help me set up a 5x5 risk acceptability matrix with P1-P5 and S1-S5 categories
- Perform a hazard analysis on software errors using Software FMEA
How to install risk-management-specialist
1Plugin marketplace
/plugin marketplace add alirezarezvani/claude-skills/plugin install ra-qm-skills@claude-code-skillsThis installs the whole ra-qm-skills plugin with all its skills
2Skills CLI (Vercel)
npx skills add alirezarezvani/claude-skills --skill risk-management-specialist -a claude-code3Manually: put the files in place
~/.claude/skills/risk-management-specialist/SKILL.mdCopy the whole skill folder; for one project use .claude/skills/risk-management-specialist/
Manually: put the files in place
.agents/skills/risk-management-specialist/SKILL.mdFor all projects use ~/.agents/skills/
Manually: put the files in place
.cursor/skills/risk-management-specialist/SKILL.mdCursor also reads .agents/skills/ and .claude/skills/
1GitHub CLI
gh skill install alirezarezvani/claude-skills risk-management-specialist2Manually: put the files in place
.github/skills/risk-management-specialist/SKILL.mdPlatform CLI
gemini skills install https://github.com/alirezarezvani/claude-skills --path ra-qm-team/skills/risk-management-specialistManually: put the files in place
.devin/skills/risk-management-specialist/SKILL.mdThe legacy .windsurf/skills/ path still works
Claude.ai
Zip the folder and upload it in Settings → Capabilities → Skills
From the same set349
The whole set →Similar
Questions about risk-management-specialist
How to install risk-management-specialist in Claude Code?
Run npx skills add alirezarezvani/claude-skills --skill risk-management-specialist -a claude-code or place the file directly at ~/.claude/skills/risk-management-specialist/SKILL.md.
What does risk-management-specialist do?
It guides medical device risk assessments following ISO 14971:2019. The skill walks through hazard classification, evaluation matrices, and control verification.
Why does this skill reject ALARP for EU MDR devices?
EU MDR Annex I mandates reducing risks as far as possible (AFAP) without economic trade-offs. Notified bodies flag ALARP because EN ISO 14971:2019/A11:2021 deviations explicitly exclude cost-benefit balancing.
Source text
SKILL.md · MIT
---
name: "risk-management-specialist"
description: Medical device risk management specialist implementing ISO 14971 throughout product lifecycle. Provides risk analysis, risk evaluation, risk control, and post-production information analysis. Use when user mentions risk management, ISO 14971, risk analysis, FMEA, fault tree analysis, hazard identification, risk control, risk matrix, benefit-risk analysis, residual risk, risk acceptability, or post-market risk.
---
# Risk Management Specialist
ISO 14971:2019 risk management implementation throughout the medical device lifecycle.
---
## Table of Contents
- [Risk Management Planning Workflow](#risk-management-planning-workflow)
- [Risk Analysis Workflow](#risk-analysis-workflow)
- [Risk Evaluation Workflow](#risk-evaluation-workflow)
- [Risk Control Workflow](#risk-control-workflow)
- [Post-Production Risk Management](#post-production-risk-management)
- [Risk Assessment Templates](#risk-assessment-templates)
- [Decision Frameworks](#decision-frameworks)
- [Tools and References](#tools-and-references)
---
## Risk Management Planning Workflow
Establish risk management process per ISO 14971.
### Workflow: Create Risk Management Plan
1. Define scope of risk management activities:
- Medical device identification
- Lifecycle stages covered
- Applicable standards and regulations
2. Establish risk acceptability criteria:
- Define probability categories (P1-P5)
- Define severity categories (S1-S5)
- Create risk matrix with acceptance thresholds
3. Assign responsibilities:
- Risk management lead
- Subject matter experts
- Approval authorities
4. Define verification activities:
- Methods for control verification
- Acceptance criteria
5. Plan production and post-production activities:
- Information sources
- Review triggers
- Update procedures
6. Obtain plan approval
7. Establish risk management file
8. **Validation:** Plan approved; acceptability criteria defined; responsibilities assigned; file established
### Risk Management Plan Content
| Section | Content | Evidence |
|---------|---------|----------|
| Scope | Device and lifecycle coverage | Scope statement |
| Criteria | Risk acceptability matrix | Risk matrix document |
| Responsibilities | Roles and authorities | RACI chart |
| Verification | Methods and acceptance | Verification plan |
| Production/Post-Production | Monitoring activities | Surveillance plan |
### Risk Acceptability Matrix (5x5)
| Probability \ Severity | Negligible | Minor | Serious | Critical | Catastrophic |
|------------------------|------------|-------|---------|----------|--------------|
| **Frequent (P5)** | Medium | High | High | Unacceptable | Unacceptable |
| **Probable (P4)** | Medium | Medium | High | High | Unacceptable |
| **Occasional (P3)** | Low | Medium | Medium | High | High |
| **Remote (P2)** | Low | Low | Medium | Medium | High |
| **Improbable (P1)** | Low | Low | Low | Medium | Medium |
### Risk Level Actions
| Level | Acceptable | Action Required |
|-------|------------|-----------------|
| Low | Yes | Document and accept; still reduce as far as possible (EU MDR) |
| Medium | After reduction AFAP | Reduce as far as possible; document why further reduction is impossible |
| High | After reduction AFAP | Reduction required; demonstrate all further options exhausted |
| Unacceptable | No | Design change mandatory |
> **EU MDR — AFAP, not ALARP:** For CE-marked devices, risks must be reduced **as far as possible (AFAP)** without economic considerations (MDR Annex I, GSPR 1–4; EN ISO 14971:2019/A11:2021 Z-annexes deviation). ALARP ("as low as reasonably practicable"), which permits cost-benefit weighing in acceptability decisions, is **not an acceptable criterion under the EU MDR** — a notified body will flag it. ISO 14971:2019 itself removed ALARP from the normative text. ALARP may persist in some non-EU jurisdictions (e.g., the UK HSE tradition); if used outside the EU, flag the deviation from EU requirements explicitly.
---
## Risk Analysis Workflow
Identify hazards and estimate risks systematically.
### Workflow: Conduct Risk Analysis
1. Define intended use and reasonably foreseeable misuse:
- Medical indication
- Patient population
- User population
- Use environment
2. Select analysis method(s):
- FMEA for component/function analysis
- FTA for system-level analysis
- HAZOP for process deviations
- Use Error Analysis for user interaction
3. Identify hazards by category:
- Energy hazards (electrical, mechanical, thermal)
- Biological hazards (bioburden, biocompatibility)
- Chemical hazards (residues, leachables)
- Operational hazards (software, use errors)
4. Determine hazardous situations:
- Sequence of events
- Foreseeable misuse scenarios
- Single fault conditions
5. Estimate probability of harm (P1-P5)
6. Estimate severity of harm (S1-S5)
7. Document in hazard analysis worksheet
8. **Validation:** All hazard categories addressed; all hazards documented; probability and severity assigned
### Hazard Categories Checklist
| Category | Examples | Analyzed |
|----------|----------|----------|
| Electrical | Shock, burns, interference | ☐ |
| Mechanical | Crushing, cutting, entrapment | ☐ |
| Thermal | Burns, tissue damage | ☐ |
| Radiation | Ionizing, non-ionizing | ☐ |
| Biological | Infection, biocompatibility | ☐ |
| Chemical | Toxicity, irritation | ☐ |
| Software | Incorrect output, timing | ☐ |
| Use Error | Misuse, perception, cognition | ☐ |
| Environment | EMC, mechanical stress | ☐ |
### Analysis Method Selection
| Situation | Recommended Method |
|-----------|-------------------|
| Component failures | FMEA |
| System-level failure | FTA |
| Process deviations | HAZOP |
| User interaction | Use Error Analysis |
| Software behavior | Software FMEA |
| Early design phase | PHA |
### Probability Criteria
| Level | Name | Description | Frequency |
|-------|------|-------------|-----------|
| P5 | Frequent | Expected to occur | >10⁻³ |
| P4 | Probable | Likely to occur | 10⁻³ to 10⁻⁴ |
| P3 | Occasional | May occur | 10⁻⁴ to 10⁻⁵ |
| P2 | Remote | Unlikely | 10⁻⁵ to 10⁻⁶ |
| P1 | Improbable | Very unlikely | <10⁻⁶ |
### Severity Criteria
| Level | Name | Description | Harm |
|-------|------|-------------|------|
| S5 | Catastrophic | Death | Death |
| S4 | Critical | Permanent impairment | Irreversible injury |
| S3 | Serious | Injury requiring intervention | Reversible injury |
| S2 | Minor | Temporary discomfort | No treatment needed |
| S1 | Negligible | Inconvenience | No injury |
See: [references/risk-analysis-methods.md](references/risk-analysis-methods.md)
---
## Risk Evaluation Workflow
Evaluate risks against acceptability criteria.
### Workflow: Evaluate Identified Risks
1. Calculate initial risk level from probability × severity
2. Compare to risk acceptability criteria
3. For each risk, determine:
- Acceptable: Document and accept (EU MDR: still reduce as far as possible)
- Reduction required (AFAP): Proceed to risk control
- Unacceptable: Mandatory risk control
4. Document evaluation rationale
5. Identify risks requiring benefit-risk analysis
6. Complete benefit-risk analysis if applicable
7. Compile risk evaluation summary
8. **Validation:** All risks evaluated; acceptability determined; rationale documented
### Risk Evaluation Decision Tree
```
Risk Estimated
│
▼
Apply Acceptability Criteria
│
├── Low Risk ──────────► Accept and document
│
├── Medium Risk ───────► Reduce as far as possible (AFAP)
│ │ Document why further reduction impossible
│ ▼
│ Further reduction possible?
│ │
│ Yes──► Implement control
│ No───► Document AFAP rationale (no economic considerations)
│
├── High Risk ─────────► Risk reduction required
│ │ Must demonstrate reduction AFAP
│ ▼
│ Implement control
│ Verify residual risk
│
└── Unacceptable ──────► Design change mandatory
Cannot proceed without control
```
### AFAP Demonstration Requirements (EU MDR)
| Criterion | Evidence Required |
|-----------|-------------------|
| All control options considered | Analysis of every feasible control per the hierarchy (design, protective measures, information) |
| Further reduction impossible | Evidence each remaining option is technically infeasible or does not further reduce risk |
| State of the art | Comparison to similar devices and current standards |
| Stakeholder input | Clinical/user perspectives |
> Economic considerations (cost of further risk reduction) **must not** enter the EU acceptability decision (MDR Annex I GSPR 2; EN ISO 14971:2019/A11:2021). Cost may inform business decisions about whether to market the device — never whether a risk is acceptable.
### Benefit-Risk Analysis Triggers
| Situation | Benefit-Risk Required |
|-----------|----------------------|
| Residual risk remains high | Yes |
| No feasible risk reduction | Yes |
| Novel device | Yes |
| Unacceptable risk with clinical benefit | Yes |
| All risks low | No |
---
## Risk Control Workflow
Implement and verify risk control measures.
### Workflow: Implement Risk Controls
1. Identify risk control options:
- Inherent safety by design (Priority 1)
- Protective measures in device (Priority 2)
- Information for safety (Priority 3)
2. Select optimal control following hierarchy
3. Analyze control for new hazards introduced
4. Document control in design requirements
5. Implement control in design
6. Develop verification protocol
7. Execute verification and document results
8. Evaluate residual risk with control in place
9. **Validation:** Control implemented; verification passed; residual risk acceptable; no unaddressed new hazards
### Risk Control Hierarchy
| Priority | Control Type | Examples | Effectiveness |
|----------|--------------|----------|---------------|
| 1 | Inherent Safety | Eliminate hazard, fail-safe design | Highest |
| 2 | Protective Measures | Guards, alarms, automatic shutdown | High |
| 3 | Information | Warnings, training, IFU | Lower |
### Risk Control Option Analysis Template
```
RISK CONTROL OPTION ANALYSIS
Hazard ID: H-[XXX]
Hazard: [Description]
Initial Risk: P[X] × S[X] = [Level]
OPTIONS CONSIDERED:
| Option | Control Type | New Hazards | Feasibility | Selected |
|--------|--------------|-------------|-------------|----------|
| 1 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |
| 2 | [Type] | [Yes/No] | [H/M/L] | [Yes/No] |
SELECTED CONTROL: Option [X]
Rationale: [Justification for selection]
IMPLEMENTATION:
- Requirement: [REQ-XXX]
- Design Document: [Reference]
VERIFICATION:
- Method: [Test/Analysis/Review]
- Protocol: [Reference]
- Acceptance Criteria: [Criteria]
```
### Risk Control Verification Methods
| Method | When to Use | Evidence |
|--------|-------------|----------|
| Test | Quantifiable performance | Test report |
| Inspection | Physical presence | Inspection record |
| Analysis | Design calculation | Analysis report |
| Review | Documentation check | Review record |
### Residual Risk Evaluation
| After Control | Action |
|---------------|--------|
| Acceptable | Document, proceed |
| Reduced AFAP | Document rationale (no economic considerations), proceed |
| Still unacceptable | Additional control or design change |
| New hazard introduced | Analyze and control new hazard |
---
## Post-Production Risk Management
Monitor and update risk management throughout product lifecycle.
### Workflow: Post-Production Risk Monitoring
1. Identify information sources:
- Customer complaints
- Service reports
- Vigilance/adverse events
- Literature monitoring
- Clinical studies
2. Establish collection procedures
3. Define review triggers:
- New hazard identified
- Increased frequency of known hazard
- Serious incident
- Regulatory feedback
4. Analyze incoming information for risk relevance
5. Update risk management file as needed
6. Communicate significant findings
7. Conduct periodic risk management review
8. **Validation:** Information sources monitored; file current; reviews completed per schedule
### Information Sources
| Source | Information Type | Review Frequency |
|--------|------------------|------------------|
| Complaints | Use issues, failures | Continuous |
| Service | Field failures, repairs | Monthly |
| Vigilance | Serious incidents | Immediate |
| Literature | Similar device issues | Quarterly |
| Regulatory | Authority feedback | As received |
| Clinical | PMCF data | Per plan |
### Risk Management File Update Triggers
| Trigger | Response Time | Action |
|---------|---------------|--------|
| Serious incident | Immediate | Full risk review |
| New hazard identified | 30 days | Risk analysis update |
| Trend increase | 60 days | Trend analysis |
| Design change | Before implementation | Impact assessment |
| Standards update | Per transition period | Gap analysis |
### Periodic Review Requirements
| Review Element | Frequency |
|----------------|-----------|
| Risk management file completeness | Annual |
| Risk control effectiveness | Annual |
| Post-market information analysis | Quarterly |
| Risk-benefit conclusions | Annual or on new data |
---
## Risk Assessment Templates
→ See references/risk-assessment-templates.md for details
## Decision Frameworks
### Risk Control Selection
```
What is the risk level?
│
├── Unacceptable ──► Can hazard be eliminated?
│ │
│ Yes─┴─No
│ │ │
│ ▼ ▼
│ Eliminate Can protective
│ hazard measure reduce?
│ │
│ Yes─┴─No
│ │ │
│ ▼ ▼
│ Add Add warning
│ protection + training
│
└── High/Medium ──► Apply hierarchy
starting at Level 1
```
### New Hazard Analysis
| Question | If Yes | If No |
|----------|--------|-------|
| Does control introduce new hazard? | Analyze new hazard | Proceed |
| Is new risk higher than original? | Reject control option | Acceptable trade-off |
| Can new hazard be controlled? | Add control | Reject control option |
### Risk Acceptability Decision
| Condition | Decision |
|-----------|----------|
| All risks Low | Acceptable |
| Medium risks reduced AFAP | Acceptable |
| High risks reduced AFAP, documented | Acceptable if benefits outweigh |
| Any Unacceptable residual | Not acceptable - redesign |
---
## Tools and References
### Scripts
| Tool | Purpose | Usage |
|------|---------|-------|
| [risk_matrix_calculator.py](scripts/risk_matrix_calculator.py) | Calculate risk levels and FMEA RPN | `python risk_matrix_calculator.py --help` |
**Risk Matrix Calculator Features:**
- ISO 14971 5x5 risk matrix calculation
- FMEA RPN (Risk Priority Number) calculation
- Interactive mode for guided assessment
- Display risk criteria definitions
- JSON output for integration
### References
| Document | Content |
|----------|---------|
| [iso14971-implementation-guide.md](references/iso14971-implementation-guide.md) | Complete ISO 14971:2019 implementation with templates |
| [risk-analysis-methods.md](references/risk-analysis-methods.md) | FMEA, FTA, HAZOP, Use Error Analysis methods |
### Quick Reference: ISO 14971 Process
| Stage | Key Activities | Output |
|-------|----------------|--------|
| Planning | Define scope, criteria, responsibilities | Risk Management Plan |
| Analysis | Identify hazards, estimate risk | Hazard Analysis |
| Evaluation | Compare to criteria, AFAP assessment (EU) | Risk Evaluation |
| Control | Implement hierarchy, verify | Risk Control Records |
| Residual | Overall assessment, benefit-risk | Risk Management Report |
| Production | Monitor, review, update | Updated RM File |
---
## Related Skills
| Skill | Integration Point |
|-------|-------------------|
| [quality-manager-qms-iso13485](../quality-manager-qms-iso13485/) | QMS integration |
| [capa-officer](../capa-officer/) | Risk-based CAPA |
| [regulatory-affairs-head](../regulatory-affairs-head/) | Regulatory submissions |
| [quality-documentation-manager](../quality-documentation-manager/) | Risk file management |